Tripp Lite series SmartPro 230V 1kVA 750W Medical-Grade Line-Interactive Tower UPS with 6 Outlets, Full Isolation, Expandable Runtime
$1,371.78
Tripp Lite series SmartPro 230V 1kVA 750W Medical-Grade Line-Interactive Tower UPS with 6 Outlets, Full Isolation, Expandable Runtime
*Items typically ship within 3-4 weeks if stock at factory
Quick Specs
- Supports a half load up to 27.5 minutes; full load up to 11 minutes
- Six IEC320 C13 hospital-grade outlets and IEC320 C14 hospital-grade plug
- Compatible with Eaton Tripp Lite power cords to simplify deployment and enhance reliability
- Leakage current less than 100 microamps
- 340 joule surge suppression rating protects connected equipment
- Alarm mute/alarm quiet control is ideal for use around patients
- Approved to EN60601-1-2:2015, 4th Edition
Overview
The SMX1200XLHG medical-grade UPS system provides dependable battery backup and is fully compliant with UL 60601-1 regulations when plugged into a hospital-grade outlet, making it ideal for use in patient care vicinities. The 1kVA/750W capacity provides up to 27.5 minutes of runtime for a half load or 11 minutes for a full load, so you can work through short power failures and perform an orderly system shutdown in a prolonged blackout.Built-in line-interactive voltage regulation corrects brownouts as low as 155V and overvoltages as high as 274V to normal 230V nominal output, without drawing battery power. The SMX1200XLHG’s built-in isolation transformer with Faraday shield offers full isolation and removes common mode noise while maintaining full AC surge suppression in all modes. Cumulative leakage current of all connected equipment is less than 100 microamps.This UPS system features six IEC320 C13 hospital-grade outlets and has a hospital-grade IEC320 C14 plug on a 15-centimeter (six-inch) dongle. All six outlets provide both backup protection and surge protection, rated to 340 joules.The SMX1200XLHG also features diagnostic LEDs, as well as a USB port and a DB9 serial port that work with included PowerAlert® software for safe, unattended system shutdown in the event of a prolonged power failure.This UPS contains a lead acid battery designed to be connected continuously to an AC power source, except during brief power outages. It is not intended for frequent charge/discharge cycles, such as mobile cart applications where the UPS is often disconnected from the power source. Using the SMX1200XLHG in this type of application will void the battery warranty.
Features
Provides Ideal UPS Backup in Healthcare Facilities
- Compliant with UL 60601-1 standards for use in patient-care vicinities, when plugged into a hospital-grade outlet
- Built-in isolation transformer with Faraday shield offers full isolation and removes common mode noise
Battery Backup and Full Surge Protection
- 27.5 minutes runtime for half load (375W); 11 minutes for full load (750W)
- All six hospital-grade outlets provide surge protection rated to 340 joules
- EMI/RFI noise filtering ensures equipment runs smoothly without line noise interference
- Resistant to electrostatic discharge
LEDs and Audible Alarms Keep You Aware of Power Status
- Indicator lights on the front of the unit show you when the UPS is on and supplying utility power, when operating in voltage correction mode or in battery mode, relative load level and low-battery warning
- Audible alarms alert you to power failure, overload and low battery conditions; the alarm’s mute/quiet mode ensures patients are not disturbed
Saves Your Files, Even When You’re Away
- Works with PowerAlert software (included) to enable safe, unattended system shutdown and files saves during prolonged power failures
- Connects to your computer via either the DB9 port or USB port
Package Includes
- SMX1200XLHG 1kVA/750W Medical-Grade Tower UPS System
- USB monitoring cable, 1.8 m (6 ft.)
- DB9 serial monitoring cable, 1.8 m (6 ft.)
- PowerAlert software
- Owner’s Manual
| Weight | 48.8 lbs |
|---|---|
| Dimensions | 13.38 × 12.2 × 18.5 in |
| UPC Code | 037332140425 |
| Cord Length | 0.66 |
| Max Input Amps | 5 |
| Output (W) | 750 |
| Output VA | 1000 |
| Joule Rating | 340 |
| Brand | |
| UPS Type | Line-Interactive |
| Input Phase | Single-Phase |
| Rated input current (Maximum Load) | 5A |
| Nominal Input Voltage(s) Supported | 230V AC |
| UPS Input Connection Type | C14 inlet |
| UPS Input Cord Length (ft.) | 0.50 |
| UPS Input Cord Length (m) | 0.2 |
| Recommended Electrical Service | 10A 230V |
| Voltage Compatibility (VAC) | 230 |
| Output Capacity (kVA) | 1 |
| Output Capacity (kW) | 0.75 |
| Nominal Voltage Details | 230v nominal inverter output voltage |
| Frequency Compatibility | 50 / 60 Hz |
| Frequency Compatibility Details | Automatic frequency selection |
| Output Voltage Regulation (Line Mode) | -18%, +8% |
| Output Receptacle Details | Compatible with Eaton Tripp Lite Power Cords |
| Output AC Waveform (Battery Mode) | PWM sine wave |
| Nominal Output Voltage(s) Supported | 220V; 230V; 240V |
| Output Receptacles | (6) C13 |
| Battery Type | Valve Regulated Lead Acid (VRLA) |
| Runtime Full Load (min.) | 11 min. (750W) |
| Runtime Half Load (min.) | 27.5 min. (375W) |
| Expandable Runtime | Yes |
| Expandable Runtime Description | Supports extended runtime with optional BP36V15-2U (limit 1) and BP36V42-3U external battery pack (multi-pack compatible) |
| External Battery Pack Compatibility | BP36V15-2U (limit 1); BP36V42-3U (multi-pack compatible); BP36V27-2US (multi-pack compatible) |
| DC System Voltage (VDC) | 36 |
| Battery Recharge Rate (Included Batteries) | Less than 4.5 hours from 10% to 90% |
| Battery Access | Battery access door allows hot swap battery replacement without powering connected equipment off |
| Internal UPS Replacement Battery Cartridge | RBC36SLT |
| Typical Battery Lifespan | 3-6 years, depending on usage |
| Voltage Regulation Description | Line interactive UPS offers 2 boost levels of brownout correction and a single level of overvoltage adjustment to maintain usable, computer grade output without consuming battery power over an input voltage range of 155 to 274V AC. |
| Overvoltage Correction | Input voltages between 245 and 274V AC are reduced by 9% |
| Undervoltage Correction | Input voltages between 183 and 199 are boosted by 10% |
| Severe Undervoltage Correction | Input voltages between 156 and 182 are boosted by 21% |
| Switches | Yes, includes 2 front panel mounted push-button switches for system enable, self-test and alarm cancel functions |
| Audible Alarm | Audible Alarm is factory muted for silent operation; Audible alarm can be restored to indicate power failure, overload conditions and Low Battery events using our Free Power Alert UPS management software. |
| LED Indicators | 5 LEDs indicate line power, battery power, overload, voltage regulation and battery low/replace status |
| EMI / RFI AC Noise Suppression | Yes |
| AC Suppression Joule Rating | 340 |
| AC Suppression Response Time | Instantaneous |
| Primary Form Factor | Tower |
| Rack Height | 0U |
| Cooling Method | UPS cooling via rear-exit fan |
| Installation Form Factors Supported with Optional Accessories | Wallmount ( UPSWM – tower mounting kit) |
| Primary UPS Depth (mm) | 231 |
| Primary UPS Height (mm) | 323 |
| Primary UPS Width (mm) | 191 |
| Shipping Dimensions (hwd / cm) | 46.99 x 30.99 x 33.99 |
| Shipping Weight (kg) | 22.14 |
| UPS Housing Material | PVC |
| UPS Power Module Dimensions (hwd, cm) | 32.26 x 19.05 x 23.11 |
| UPS Power Module Weight (kg) | 20.64 |
| Unit Weight (kg) | 20.64 |
| Operating Temperature Range | 32° to 104°F (0° to 40°C) |
| Storage Temperature Range | 5° to 122°F (-15° to 50°C) |
| Relative Humidity | 0-95% non-condensing |
| AC Mode BTU / Hr. (Full Load) | 252.2 |
| Battery Mode Heat Dissipation (BTU/Hr) @ Full Load | 252.2 |
| Network Management Card Description | Network management not supported |
| Communications Interface | USB; DB9 Serial |
| Transfer Time | 2-4 milliseconds |
| Low Voltage Transfer to Battery Power (Setpoint) | 155 |
| High Voltage Transfer to Battery Power (Setpoint) | 274 |
| Cold Start (Startup in Battery Mode During a Power Failure) | Cold-start operation supported |
| High Availability UPS Features | Surge/noise protection |
| Grounding Details | Ground Potential Equalization Connector: POAG-S6/20 |
| UPS Applications | Hospital/Medical |
| Product Certifications | UL 60601-1 |
| Product Compliance | RoHS; CE (Europe) |
| Product Warranty Period (International) | 2-year limited warranty |
Compatible accessories
-
IN STOCKTripp Lite series External 36V 3U Rack/Tower Battery Pack for Select UPS Systems (BP36V42-3U), TAA
Manufacturer: Tripp LitePart Number: BP36V42-3U $2,520.41 -
IN STOCKTripp Lite series External 36V Battery Pack for Select UPS Systems, 2U Rack/Tower
Manufacturer: Tripp LitePart Number: BP36V27-2US $1,268.53 -

Tripp Lite series External 36V 2U Rack/Tower Battery Pack for Select UPS Systems (BP36V15-2U)
Manufacturer: Tripp LitePart Number: BP36V15-2U $1,007.48









Reviews